History Data
History Data
Gauge Calibration Control Record (group#159)
Approved
Checked
Gauge no. IP-ID-AF-3.00- 900J E
Gauge control name Inside Go-No go gauge Ø3.00 900J
Approved
Checked
Tool standard name Accuracy Resolution Temp. control
MICRO METER / C-23 ±1 um. 0.001 mm. 18 -23º C
Calib. date Next Calib. Cycle Calibrated by
2026-09-29 10:25 AM 2027-03-29 6 mth NUENGRUTHAI
Point
1
Measurement sub group
1st 2nd 3rd
Spec. 3.000 3.000 3.000
±Tol. 0.000 / -0.010 0.000 / -0.010 0.000 / -0.010
Actual 2.998 2.998 2.999
Diff.
-0.002
-0.002
-0.001
Judge
Ok
Ok
Ok
Point
2
Measurement sub group
1st 2nd 3rd
Spec. 2.900 2.900 2.900
±Tol. 0.010 / 0.000 0.010 / 0.000 0.010 / 0.000
Actual 2.901 2.903 2.903
Diff.
0.001
0.003
0.003
Judge
Ok
Ok
Ok
Factory Data
  • Wanna join ?
  • . .
Small Tools
UICT Factory Enterprise Edition