Gauge Calibration Control Record (group#f6a)
Approved
Checked
Gauge no. IDG-3.0-AF BI
Gauge control name ID Tube After Auto Frettage Type UPH-27
Approved
Checked
Tool standard name Accuracy Resolution Temp. control
MICRO METER / C-23 ±1 um. 0.001 mm. 18 -23º C
Calib. date Next Calib. Cycle Calibrated by
2026-10-06 01:44 PM 2027-04-06 6 mth NUENGRUTHAI
Point
1
Measurement sub group
1st 2nd 3rd
Spec. 3.050 3.050 3.050
±Tol. 0.000 / -0.010 0.000 / -0.010 0.000 / -0.010
Actual 3.043 3.043 3.047
Diff.
-0.007
-0.007
-0.003
Judge
Ok
Ok
Ok
Point
2
Measurement sub group
1st 2nd 3rd
Spec. 2.900 2.900 2.900
±Tol. 0.010 / 0.000 0.010 / 0.000 0.010 / 0.000
Actual 2.902 2.902 2.905
Diff.
0.002
0.002
0.005
Judge
Ok
Ok
Ok