History Data
History Data
Gauge Calibration Control Record (group#15f)
Approved
Checked
Gauge no. SAIP-PG-DI8QD1 J
Gauge control name Plug gauge SAIP-PG-DI8QD1
Approved
Checked
Tool standard name Accuracy Resolution Temp. control
MICRO METER / C-23 ±1 um. 0.001 mm. 18 -23º C
Calib. date Next Calib. Cycle Calibrated by
2026-09-27 01:29 PM 2027-03-27 6 mth NUENGRUTHAI
Point
1
Measurement sub group
1st 2nd 3rd
Spec. 4.800 4.800 4.800
±Tol. 0.000 / -0.050 0.000 / -0.050 0.000 / -0.050
Actual 4.788 4.787 4.786
Diff.
-0.012
-0.013
-0.014
Judge
Ok
Ok
Ok
Point
2
Measurement sub group
1st 2nd 3rd
Spec. 4.500 4.500 4.500
±Tol. 0.050 / 0.000 0.050 / 0.000 0.050 / 0.000
Actual 4.538 4.541 4.539
Diff.
0.038
0.041
0.039
Judge
Ok
Ok
Ok
Factory Data
  • Wanna join ?
  • . .
Small Tools
UICT Factory Enterprise Edition